4-electrode sensors 546
Description
discover more
El sensor de conductividad de cuatro electrodos 546 se puede utilizar con el mismo hardware y accesorios que los sensores de pH/ORP_cc781905-5cde-3194- bb3b-136bad5cf58d_546. Ideal for conductivity measurement in clean chemical applications, the 546 sensor provides high accuracy in a much smaller package than competitive toroidal sensor technology.
- 3/4 inch male NPT process connection
- Conductivity measurements up to 2 Siemens (depending on cell constant)
- Kynar body and PEEK insulation
- Autocompensation for corrosion and scale buildup (depends on analyzer)
- Insertion depths from 0.5 to 5 inches
- Electrode options in 316 stainless steel, titanium and Hastelloy
As the name suggests, four-electrode sensors add an additional pair of electrodes to the two-electrode sensor design. This second pair of electrodes provides sensor diagnostics that can then be used to compensate the measurement if scale or particles accumulate on the electrodes. Four-electrode conductivity sensors can resist coating and scale that could otherwise foul a traditional two-electrode sensor._cc781905-5cde -3194-bb3b-136bad5cf58d_Typical applications include the following:
- leak detection
- condensate return
- Salinity
- chemical concentration
- on-site cleaning
Four electrode sensor designs maintain a constant current through two of the electrodes and allow the drive voltage to change. If fouling occurs, the drive voltage can be increased to compensate for the measurement.